Equilibrate reagents and prepare standards and samples
Bring kit components to room temperature, then prepare the serial standard and sample inputs for consistent binding.
- STEP 1Equilibrate all reagents to room temperature30 min
Remove kit components from 4°C storage and allow to equilibrate at room temperature (20–25°C) for a minimum of 30 minutes. Prepare 1× Wash Buffer: dilute the 10× concentrate 1:10 with ddH₂O. Mix thoroughly. Calculate volume needed: 300 µL/well × 4 washes × number of wells used (add 10–20% overage).
WhyCold reagents reduce antibody-binding kinetics and enzyme activity, leading to underestimated signals and a compressed standard curve. Uniform temperature across all wells is critical for intra-plate reproducibility (Tighe et al., 2015).
- STEP 2Reconstitute and serially dilute the standard30 min
Reconstitute lyophilized standard: add the volume of Sample Diluent specified by the kit. Allow to sit undisturbed for 15 minutes; mix by gentle inversion — do not vortex. Label 7 microcentrifuge tubes (S6 through S1) plus one tube for the blank (S0). Pipette Sample Diluent into tubes S6–S1 (volume per kit; commonly 300 µL each). Transfer the volume specified by the kit from the reconstituted top standard (S7) into tube S6; mix by pipetting up and down 10 times. Change tips. Perform 2-fold serial dilution: transfer from S6 → S5 → S4 → S3 → S2 → S1, changing tips between each transfer. S0 = Sample Diluent only (0 pg/mL blank). Do not add standard. Run all standards in duplicate.
WhySerial dilution accuracy is the single largest source of standard curve error. Changing tips between transfers prevents carryover; consistent pipetting motion (aspirate and dispense 10×) ensures complete mixing at each step. A 2-fold dilution series is standard for most cytokine ELISA kits and produces a 7-point calibration range spanning 2–3 orders of magnitude.
CautionDo not vortex the reconstituted standard — protein aggregation can cause non-linearity at the high end of the curve.
- STEP 3Prepare samples20 min
Cell culture supernatant: centrifuge at 1,000 × g for 10 min at 4°C to pellet debris. Collect the clarified supernatant. Use immediately or aliquot and store at −80°C; avoid more than 2 freeze–thaw cycles. Serum: allow blood to clot at room temperature for 30 min, then centrifuge at 2,000 × g for 10 min. Collect serum layer (top). Plasma: collect blood into EDTA or heparin tube; centrifuge at 2,000 × g for 10 min. Collect the plasma layer (top). Dilute samples in Sample Diluent as required (start with a 2-fold dilution if analyte concentration is unknown; dilutions must fall within the linear range of the standard curve).
WhyCellular debris and lipid in undiluted samples can cause matrix interference (hook effect at high analyte concentrations, non-specific binding). Serum is preferred over plasma for cytokine ELISA when anticoagulants may interfere with the assay (Lequin, 2005).